Creosote Oil Standard on Rxi-1HT (15 m x 0.25 mm x 0.10 µm)

image
GC_EV1367
ColumnRxi®-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleCreosote oil standard (cat.# 31838)
Diluent:Methylene chloride
Conc.:50,000 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 250:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:437.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 20 °C/min (hold 1.9 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC® flow calculator (http://www.restek.com/ezgc-mtfc).
Voir en PDF

RECHERCHES SIMILAIRES

Total Petroleum Hydrocarbons

,

Total Recoverable Petroleum Hydrocarbons

,

Alkanes

,

Aromatics

,

Creosote Oil

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.