GCxGC-TOFMS Analysis of Multi-Pesticide Standard Using Rxi-5Sil MS and Rtx-200

GC_FF1241
Column | Rxi®-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Rtx®-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15124) | |
Sample | Custom pesticide standard |
Diluent: | Acetonitrile |
Conc.: | 500 pg/µL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1.0 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.5) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | Rxi®-5Sil MS: 90 °C (hold 1.0 min) to 310 °C at 5 °C/min (hold 2.0 min) Rtx®-200: 95 °C (hold 1.0 min) to 315 °C at 5 °C/min (hold 2.0 min) |
Carrier Gas | He, corrected constant flow (2 mL/min) |
Modulation | |
Modulator Temp. Offset: | +20 °C |
Second Dimension Separation Time: | 3 sec |
Hot Pulse Time: | 0.9 sec |
Cool Time between Stages: | 0.6 sec |
Detector | MS |
Mode: | |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Ionization Mode: | EI |
Acquisition Range: | 45 to 550 amu |
Spectral Acquisition Rate: | 100 spectra/sec |
Instrument | LECO Pegasus® 4D GCxGC-TOFMS |
Notes | Rtx®-200 (cat.# 15124) is a 2 m column. A 1.3 m section was cut off and used as the second dimension column. |
RECHERCHES SIMILAIRES
Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.