Innovative Chromatography Products
www.restek.fr
www.restek.fr

Diesel on Rxi-1ms and Rtx-FLD (GCxGC-TOFMS, TIC)

GC_PC1203
Column | Rxi®-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326) with Rxi®-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326) |
Rtx®-FLD 1.5 m, 0.25 mm ID, 0.10 µm, | |
Sample | Diesel fuel #2 composite (cat.# 31093) |
Diluent: | Methylene chloride |
Conc.: | 5,000 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4 mm Precision® inlet liner w/wool (cat.# 23305.1) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 100 mL/min |
Oven | |
Oven Temp.: | Rxi®-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min) Rtx®-FLD: 45 °C (hold 1 min) to 325 °C at 1.3 °C/min (hold 1 min) |
Carrier Gas | He, corrected constant flow (2 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 3 sec |
Hot Pulse Time: | 1 sec |
Cool Time between Stages: | 0.5 sec |
Detector | TOFMS |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | TOF |
Source Temp.: | 250 °C |
Electron Energy: | 70 eV |
Mass Defect: | 50 mu/100 u |
Solvent Delay Time: | 10 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45 to 450 amu |
Spectral Acquisition Rate: | 200 spectra/sec |
Instrument | LECO Pegasus® 4D GCxGC-TOFMS |
Notes | The Rtx®-FLD column is an experimental column and is not yet commercially available. |
RECHERCHES SIMILAIRES
Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.